歡迎來到匠仁醫療設備有限公司網站!
內窺鏡系統及周邊設備維修范圍:
Scope of maintenance for endoscope system and peripheral equipment:
軟鏡主機、軟鏡光源、超聲乳化手柄、電子鏡、纖維鏡、攝像系統、攝像頭、氣腹機
Soft mirror host, soft mirror light source, ultrasonic emulsification handle, electronic mirror, fiber mirror, camera system, camera, pneumoperitoneum machine
高頻電刀、冷光源、超聲刀、動力系統、導光束、手術器械等
High frequency electric knife, cold light source, ultrasonic knife, power system, guided beam, surgical instruments, etc
醫療用內窺鏡主要性能參數
Main performance parameters of medical endoscopes
(4)機械性能:
(4) Mechanical performance:
①密封性:內窺鏡的密封性關系到成像質量,如密封不好就容易滲水,破壞光學系統,影響觀察。通水閥鏡鞘與膀胱鏡的錐體配合處應密合,在1min內滲水應不超過5滴,否則配合不好容易漏水影響手術。
① Sealing: The sealing of the endoscope is related to the imaging quality. If the sealing is not good, it is easy to leak water, damage the optical system, and affect observation. The fitting between the sheath of the water valve and the cone of the cystoscope should be tight, and the water leakage should not exceed 5 drops within 1 minute. Otherwise, poor fitting may cause water leakage and affect the surgery.
②表面粗糙度
② Surface roughness
③連接部位牢固性
③ Durability of connection parts
(5)絕緣性能:
(5) Insulation performance:
主要是電子內窺鏡,相關電氣隔離部位的絕緣結構,如與CCD攝像頭相配接的目鏡罩、冷光源接口、導光索等相關附件的絕緣性能。
Mainly the insulation structure of electronic endoscopes and related electrical isolation parts, such as the insulation performance of eyepiece covers, cold light source interfaces, light guide cables, and other related accessories that are matched with CCD cameras.
(6)絕熱性能:
(6) Insulation performance:
在醫用內窺鏡中,由于內窺鏡是侵入性檢查工具,為了避免內窺鏡工作的時候對人體的傷害,現在一般都采用冷光源,在光輸出口設置紅外濾光片,。大限度地限制紅外光的輸出。
In medical endoscopes, as endoscopes are invasive examination tools, in order to avoid harm to the human body during the operation of endoscopes, cold light sources are generally used, and infrared filters are installed at the light output outlet. Maximize the output of infrared light.
內窺鏡測漏流程
Endoscopic leak detection process
1、內鏡蓋好防水帽掛于主機上;
1. Cover the endoscope with a waterproof cap and hang it on the host;
2、打開光源開關;
2. Turn on the light source switch;
3、將測漏器插入主機;
3. Insert the leak detector into the host;
4、檢查測漏器壓力;
4. Check the pressure of the leak detector;
5、測漏器與鏡子相連接;
5. The leak detector is connected to the mirror;
6、充氣3秒鐘看鏡子前頭是否漏氣;
6. Inflate for 3 seconds to check if there is any air leakage in the front of the mirror;
7、將鏡子完全放入水中;
7. Completely immerse the mirror in water;
8、使各旋鈕均處于放松狀態并旋轉大小旋鈕,檢查有無漏氣(約30秒);
8. Place all knobs in a relaxed state and rotate the large and small knobs to check for air leakage (approximately 30 seconds);
9、將鏡子取出擦干;
9. Take out the mirror and dry it;
10、測漏器與主機分離(約30秒,目的是放掉鏡內的氣體);
10. The leak detector is separated from the host (for about 30 seconds, with the aim of releasing the gas inside the mirror);
11、測漏器與鏡子分離,操作完畢。
11. The leak detector is separated from the mirror and the operation is complete.
奧林巴斯纖維細鏡先端組件的組裝過程
The assembly process of Olympus fiber mirror tip components
二、清理先端頭
2、 Clean the leading end
1.用抬鉗座夾住先端頭,蓋玻璃面朝上,將1mm直徑的鋼棒對著鉗子管出口位金屬接頭,用小錘敲擊,直到把金屬接頭連著部分鉗子管往下脫出,接著用小號一字批將殘余在鉗子管安裝孔里的AB膠清理干凈;
1. Use the pliers holder to clamp the front end, with the glass cover facing upwards. Place a 1mm diameter steel rod towards the metal joint at the outlet of the pliers tube, and tap it with a small hammer until the metal joint is partially detached from the pliers tube. Then, use a small batch to clean the remaining AB glue in the installation hole of the pliers tube;
2.將合適的鋼棒插進空心的像束金屬部,用斜口鉗夾住露出先端頭外的金屬部,把金屬頭完全夾取出來。
2. Insert a suitable steel rod into the hollow bundle of metal parts, clamp the metal part exposed outside the tip with diagonal pliers, and completely remove the metal head.
3.用磨尖的小號批,在放大鏡下清理像束安裝孔壁的AB膠,(小心不能碰到物鏡,以免刮花物鏡)直到干凈后把物鏡倒出來,(有時需要用力敲擊先端頭才能倒出來)然后一手用鑷子夾住物鏡,一手用鑷子夾鏡頭紙,沾酒精把貼在物鏡上的光學膠拭擦掉,直鏡片干凈無塵。
3. Use a sharpened small batch to clean the AB adhesive on the wall of the image beam installation hole under a magnifying glass (be careful not to touch the objective lens to avoid scratching the objective lens) until it is clean. Then, pour out the objective lens (sometimes requiring strong tapping on the front end to pour it out), and use tweezers to clamp the objective lens with one hand, and use tweezers to clamp lens paper with the other hand. Wipe off the optical adhesive attached to the objective lens with alcohol until the lens is clean and dust-free.
4.用合適轉頭的手動轉把光束安裝孔里的光纖清理出來,當轉到接近底的時候放慢以免磨花光束蓋玻璃。
4. Use a suitable manual rotation head to clean the optical fibers from the installation hole of the beam. When the rotation is close to the bottom, slow down to avoid grinding the glass of the beam cover.
5.用鏡頭紙沾酒精清理像束及光束安裝孔。
5. Clean the image beam and beam mounting holes with lens paper dipped in alcohol.
本文由內窺鏡設備維修提供技術支持,更多的詳細精彩內容請點擊我們的網http://www.yojeim.com我們將會全心全意為您提供滿意的服務。
This article provides technical support for endoscopic equipment maintenance. For more detailed and exciting content, please click on our website http://www.yojeim.com We will wholeheartedly provide you with satisfactory service.
樊經理:13153199508 李經理:13873135765
公司地址:山東省濟南市槐蔭區美里東路3000號德邁國際信息產業園6號樓101-2室 湖南省長沙市雨花區勞動東路820號恒大綠洲14棟2409室