醫療用內窺鏡設備定期保養一般由儀管員配合工程技術人員完成,它是一項不斷循環進行的有組織有計劃的維修措施,這有利于掌握儀器的運行規律,有利于出現故障后的查找:定期保養的內容和時間,不同儀器有不同的做法.一般可以分3個等級:
The regular maintenance of medical endoscope equipment is generally completed by the instrument controller in cooperation with the engineering technicians. It is an organized and planned maintenance measure that is carried out continuously and circularly, which is conducive to mastering the operation rules of the instrument and finding the fault after it occurs. The content and time of regular maintenance are different for different instruments. Generally, it can be divided into three levels:
(1)一保:一般可以一個月一個季度進行一次.主要內容除廠日常維護的工作外,可以拆開機殼,各處積塵,污垢、異物、緊固螺絲、添加潤滑劑;檢查務器件、元件有無磨損、變形、燒蝕、擊穿、松動,受潮,老化、接地等情況;檢測各組電源電壓及紋波,檢查高壓部件運行和接觸情況等;
(1) First maintenance: generally, it can be carried out once a month to once a quarter. In addition to the daily maintenance work of the factory, the main contents can be disassembled, dust, dirt, foreign matter, fastening screws, and lubricant can be added everywhere; Check the service devices and components for wear, deformation, ablation, breakdown, looseness, moisture, aging, grounding and other conditions; Check the power supply voltage and ripple of each group, and check the operation and contact of high-voltage components;
(2)二保:一般可以半年一年進行一次。主要內容除做好一保外,可以對整機控制臺上的各個儀表及操作控制系統的靈敏度,精度進行測試校正和計量檢定,更換高壓發生器絕緣汕等到期的損耗品,對電路中各測試點的電壓,波形進行系統檢測和做拉偏試驗;
(2) Secondary insurance: generally, it can be carried out once a year or half a year. In addition to the first guarantee, the main contents include testing, calibration and metrological verification of the sensitivity and accuracy of each instrument and operation control system on the console of the whole machine, replacement of expired loss products such as high-voltage generator insulation, systematic detection of the voltage and waveform of each test point in the circuit, and pull bias test;
(3)三保:一般可以2-4年進行一次:主要內容除做好二保外,要寸可以將整機進行全部拆卸子以清洗檢修,超過使用期的元器件應盡量更換或修復,應對儀器進行較為徹底的調試,恢復其工作精度和性能,達到或超過新機的程度是完全有可能的。
(3) Three guarantees: generally, it can be carried out once every 2-4 years. In addition to the second guarantee, the whole machine can be fully disassembled for cleaning and maintenance. Components beyond the service life should be replaced or repaired as much as possible. The instrument should be thoroughly debugged to restore its working accuracy and performance. It is completely possible to reach or exceed the level of the new machine.
將擦干后的內鏡置于多酶洗液中浸泡,時間按使用說明。使用后立即用流動水徹底清洗,并擦干。器械的軸節部、彎曲部、管腔內用軟毛刷徹底刷洗,刷洗時注意避免劃傷鏡面。徹底清洗內鏡各部件,管腔應當用高壓水槍徹底沖洗,可拆卸部分必須拆開清洗,并用超聲清洗器清洗5~10分鐘。
Soak the dried endoscope in multi-enzyme lotion for the time specified in the instructions. Wash thoroughly with flowing water and wipe dry immediately after use. Thoroughly brush the shaft joint, bending part and lumen of the instrument with a soft brush, and pay attention to avoid scratching the mirror surface when brushing. Thoroughly clean all parts of the endoscope. The lumen should be thoroughly washed with a high-pressure water gun. The detachable part must be disassembled for cleaning and cleaned with an ultrasonic cleaner for 5 to 10 minutes.
在使用內窺鏡之前,一定要認真閱讀相關使用手冊,在了解內窺鏡的設計、制作工藝、性能等方面的情況的同時,也要注意相關使用操作說明和警告提示文字,除了要正確使用外,在清洗和存儲保養等方面也要注意。主要從內窺鏡的使用環境要求,手術前后的使用操作,以及術后的清洗方面來談談醫用光學/光纖內窺鏡清洗、維護等相關問題。有問題就來撥打我們的內窺鏡維修電話
http://www.yojeim.com吧!
Before using the endoscope, you must carefully read the relevant user manual. While understanding the design, manufacturing process, performance and other aspects of the endoscope, you should also pay attention to the relevant operating instructions and warning text. In addition to correct use, you should also pay attention to cleaning, storage and maintenance. Next, we will talk about the cleaning and maintenance of the medical optical/optical fiber endoscope from the aspects of the use environment requirements of the endoscope, the use and operation before and after the operation, and the cleaning after the operation. If you have any questions, please call our endoscope maintenance hotline http://www.yojeim.com Come on!